Sundy calorimeters can be used to determine the calorific value of coal, coke, petroleum, cement black meal, solid biomass fuels and other combustibles.
Conformance with Standards
GB/T213-2008 Standard Test Method for Calorific Value of Coal ASTM D5865-2010 Standard Test Method for Gross Calorific Value of Coal and Coke ISO1928 Solid Mineral Fuels – Determination of gross calorific value by the bomb calorimetric method and calculation of net calorific value
Specifications
Analysis Time: 15~25min
Heat capacity precision:≤0.15%
Heat capacity stability:≤0.20% within three months
Temperature resolution: 0.0001k
Power Requirement:220 V±10%, 50Hz,
Max Power: 0.5kw
Gas Requirements: Oxygen (99.5% purity)
Water Requirements: Distilled water
Vertical Type, Jacket Water
≥40kgs to ensure the stability of jacket water temperature
High automation
Automatic oxygen filler equipped, oxygen filling by one press.
Automatic sample ignition.
Automatic temperature rise measurement and result calculation.
High temperature resolution
Temperature control utilizes a PT1000 Platinum resistance probe for temperature resolution to the nearest 0.0001k to improve
the test precision.
Durable bombs need hydrostatic pressure test only once a year.
With PCI interface, several calorimeters can be controlled by a single PC.
Connected with balance and network by standard interface RS232, real time data can be transmitted through internal network.
Easy-to-use Windows-based software, easy data handling.
Sundy calorimeters can be used to determine the calorific value of coal, coke, petroleum, cement black meal, solid biomass fuels and other combustibles.
Conformance with Standards
GB/T213-2008 Standard Test Method for Calorific Value of CoalASTM D5865-2010 Standard Test Method for Gross Calorific Value of Coal and CokeISO1928 Solid Mineral Fuels – Determination of gross calorific value by the bomb calorimetric method and calculation of net calorific value
Specifications
Analysis Time:<25min
Heat capacity precision:≤0.2%
Heat capacity stability:≤0.20% within three months
Temperature resolution: 0.0001k
Power Requirement:220 V±10%, 50Hz,
Max Power: 0.5kw
Gas Requirements: Oxygen (99.5% purity)
Water Requirements: Distilled water
Features
Weighing the bucket water and measuring the water temperature manually, then the calorimeter will
finish the whole test automatically.
Simple structure, easy for maintenance.
Durable bombs need hydrostatic pressure test only once a year.
With PCI interface, several calorimeters can be controlled by a single PC.
Connected with balance and network by standard interface RS232, real time data can be transmitted
by internal network.
Easy to use Windows-based operating software, easy data handling.